GLpH - automated pH measurement

Please note: As of 2009 we have discontinued production of new GLpH systems.

This does not effect our commitment to supporting existing GLpH users or the availability of spare parts and service for existing Sirius customers.

If you have a need for automated pH measurement then contact us to discuss your requirements.

Overview

For accurate and automated measurement of pH in a wide range of samples

  • High degree of accuracy
  • Easy to use
  • Automatic calibration and reminders
  • Available in single or autosampler versions
  • One-button operation for most measurements
  • Built-in GLP
  • 'Normal' and 'Supervisor' access modes

GLpH automates all the actions of pH measurement to create the world's most accurate pH meter. The electrode, stirrer and temperature probe are moved automatically between calibration buffer solutions, three wash solutions and the sample. All the errors associated with manual pH measurement are eliminated. The accuracy and repeatability are outstanding.

Sirius GLpH is used in Quality Control laboratories and process monitoring from petrochemicals to personal care. GLP protocols are built-in. Calibration can be pre-programmed to take place at user-defined intervals, so GLpH is always calibrated and ready to go.

Back to top

GLpH video

This video shows a single sample version of GLpH measuring the pH of a sample of Honey. In the video you can see the washing steps that clean the electrode automatically before and after the pH measurement, and you can also see the collection of the pH point by real-time recording of the electrode stability. When dealing with a viscous sample, the instrument waits after exiting the sample to allow drips to fall from the electrode before moving.

Get Flash player to see this video.

Back to top

Technical specification

pH electrodes

Ag/AgCI, double junction reference, general purpose (supplied as standard)
Ag/AgCI, double junction reference, green glass (for high pH applications)
Ag/AgCI double junction reference, micro diameter 4.3mm
Ag/AgCI flat end, Teflon junction* (for plates, shallow solution depth)
Ag/AgCI sleeve style Teflon junction* (for high electrode flow rate application)

Electrode fill sol.

Standard Sirius fill solution (*Teflon junctions require gel filling solution)

Electrode storage /calibration

Two positions for each buffer solution, one wash and one calibration

Single sample version

Sample placed in position by hand

Temp. probe

PT100 allows automatic temperature compensation

Probe movement

Fully automated X-Z axis

Resolution of displayed data

0.1, 0.01, 0.001 pH units (selectable)

Calibration

pH7 (or 6.86) buffer, then pH4 and/or pH10 (or 9.18); Buffers traceable to BS, DIN or NIST

Washes

Two water washes and one special wash station provided

Storage position

One (pH7 Buffer solution)

Contamination prevention

Probe assembly never travels over unmeasured samples

Measurement methods

20 available: in automated operation it is possible to interrupt methods to introduce 'rush' samples

Stirrer

Overhead, variable speed

Autosampler version tray

Holds 32 (33mm diameter, plastic) or 50 (25mm glass) sample vials, option to be temperature controlled with external circulating bath.

Keypad

Liquid-crystal 16-line display for text and graphics, 22-buttons for numeric data entry

Keyboard

Mini keyboard for alphanumeric data entry, standard on Autosampler but optional in Single Sample

pH-range

0-14 displayed. Accurate between 1.8 and 12.2

Sample container dimensions

Max height 135mm, minimum diameter 25mm (variable if automation disabled)

Data reported

mV, Temperature, pH by time or stability

Automated

For unattended and overnight operation. Data upload. Maintenance prompts.

Services required

One mains power inlet (110 or 240 V)

Connection to the PC

Cable from GLpH to PC serial comm port

Software

GLpH Firmware Version 2.014

Dimensions of single sample

W570 x D510 x H470mm

Weight

18kg

Dimensions of packaging

W660 x D510 x H320mm

Weight of packed

21kg

Dimensions of Autosampler

W610 x D500 x H570mm

Weight of Auto/s

22kg

Dimensions of packaging

W710 x D600 x H280mm (for the autosample base only)

Weight of packed

25kg (for the autosampler only)

Back to top

Literature

Electronic copies of our GLpH brochure and technical specifications are available for download:

> GLpH Brochure
> GLpH Technical Specifications
Back to top

New Scientific Paper Published

07.12.09

New paper from Sirius explains how our solubility measurements work, and introduces new methods for evaluating supersaturation.

Read more

New Scientific Paper Published

01.07.10

New paper - Study of pH-dependent solubility of organic bases. Revisit of Henderson-Hasselbalch relationship.

Read more

ACS Fall 2010 National Meeting, Boston

22.08.10

Sirius will exhibit at this meeting.

Read more

Physchem Forum 9, Barcelona

17.09.10

Sirius will exhibit at and sponsor the next Physchem Forum meeting in Barcelona

Read more

Main site navigation
HOME
PRODUCTS
INSTRUMENTATION
SiriusT3 - NEW
GLpKa
D-PAS
PCA200
Delta-8
Electrodes
GLpH
Vinotrate
SOFTWARE
RefinementPro 2
Fast D-PAS
Solubility
GI Dissolution - NEW
ANALYTICAL SERVICE
Medicinal Chemistry &    Hit to Lead
Preformulation
Formulation and Development
Consultancy
CUSTOMER SUPPORT
Maintenance
Service
Technical support
Material Safety Datasheets
Certificates of conformity
Electrode warranty forms
SCIENCE
PhysChem
pKa
logP
Solubility
Wine chemistry
TRAINING
Training sub menu
On-line training
In house training
Training courses in USA
Training courses in the UK
LITERATURE
Literature sub menu
Brochures
Bibliography
Posters
Slides
CONTACT
ABOUT
History
News archive
Events archive